Sterilization Trays

Product Overview

2DHD Camera System Tray

The VP100X Sterilization Tray Series allows for the proper storage and reprocessing of our 2DHD autoclavable devices that can be used in multi-specialty, minimally invasive surgical procedures. These components include camera heads, scopes, sheaths, bridge systems, cannula’s and light guides.

3DHD Camera System Tray

The VP1008 sterilization tray allows for the proper storage and reprocessing of our 3DHD system components including camera heads, scopes and light guide.

Video Entended LengthTray (Radel)
Method Steam (Wrapped)
Cycle Pre-Vacuum
Temperature 270°F (132°C)
Exposure 4 Minutes
Dry Cycle 35 Minutes
Video Tray Single Level (Radel)
Method Steam (Wrapped)
Cycle Pre-Vacuum
Temperature 270°F (132°C)
Exposure 4 Minutes
Dry Cycle 40 Minutes
Video Arthroscope Tray (Radel)
Method Steam (Wrapped)
Cycle Pre-Vacuum
Temperature 270°F (132°C)
Exposure 4 Minutes
Dry Cycle 15 Minutes
Video Laparoscope Tray (Radel)
Method Steam (Wrapped)
Cycle Pre-Vacuum
Temperature 270°F (132°C)
Exposure 4 Minutes
Dry Cycle 25 Minutes
3DHD Camera System Tray
Method Gas
Temperature 131°F +/-5°F (55°C +/-3°C)
Relative Humidity 35-70%
Exposure 60 Minutes
Aeration 11 Hours
Sterilization Tray
Video Entended LengthTray (Radel) vp1003
Video Tray Single Level (Radel) vp1005
Video Arthroscope Tray (Radel) vp1006
Video Laparoscope Tray (Radel) vp1007
3DHD Camera System Tray vp1008